Specifications
Introduction
Frequency
Measurement Range
Displayed Values
Note
Note
Note
Center Frequency
Unless noted, all specifications are for AUTOCOUPLED FUNCTION
operation. Where specifications are subject to minimization with the
error-correction routine, corrected limits are given unless noted.
Nominal values provide useful, but nonwarranted, information about
functional performance.
100 Hz to 1500 MHz through two RF inputs: 100 Hz to 1500 MHz
dc coupled, and 100 kHz to 1500 MHz ac coupled.
Frequency Reference Error and Accuracy
Aging rate
Temperature stability ~7 x 10eg over 0 "C to 55 "C range (25 "C reference)
The term frequency reference error, when used later in this manual,
is defined as: f [aging rate/day x number of days since calibration +
temperature stability].
When the analyzer is in STANDBY, the frequency reference
temperature is maintained at a steady state. For more information,
see "Warm-Up Time" under "General" in this chapter.
Changes in line voltage, gravitational field, and other environmental
conditions will affect the frequency reference accuracy.
Range
0 Hz to 1500 MHz
Accuracy
Specifications 3-1