Specifications Introduction; Frequency; Measurement Range; Displayed Values - HP 8568B Installation And Verification Manual

Spectrum analyzer includes option 001, option 400, option 462, and option 857
Table of Contents

Advertisement

Specifications
Introduction

Frequency

Measurement Range

Displayed Values

Note
Note
Note

Center Frequency

Unless noted, all specifications are for AUTOCOUPLED FUNCTION
operation. Where specifications are subject to minimization with the
error-correction routine, corrected limits are given unless noted.
Nominal values provide useful, but nonwarranted, information about
functional performance.
100 Hz to 1500 MHz through two RF inputs: 100 Hz to 1500 MHz
dc coupled, and 100 kHz to 1500 MHz ac coupled.
Frequency Reference Error and Accuracy
Aging rate
Temperature stability ~7 x 10eg over 0 "C to 55 "C range (25 "C reference)
The term frequency reference error, when used later in this manual,
is defined as: f [aging rate/day x number of days since calibration +
temperature stability].
When the analyzer is in STANDBY, the frequency reference
temperature is maintained at a steady state. For more information,
see "Warm-Up Time" under "General" in this chapter.
Changes in line voltage, gravitational field, and other environmental
conditions will affect the frequency reference accuracy.

Range

0 Hz to 1500 MHz
Accuracy
Specifications 3-1