Keysight N9020A MXA Signal Analyzer
Functional Tests Guide
2
Displayed Average Noise Level (DANL)
Test Limits (with 0 dB input attenuation)
See
The Displayed Average Noise Level (DANL) of the signal analyzer is measured
across a 10 kHz frequency span at several center frequencies. The analyzer
input is terminated into a 50Ω load. A test is performed to assure the
measurement is not performed in the presence of a residual response. The
measurement is then averaged, and the result is normalized to a 1 Hz
bandwidth.
Item
Termination, 50Ω Type-N(m)
Figure 2-1
DANL Test Setup
Table 2-1
for values.
Critical Specifications
(for this test)
Frequency: DC to 18 GHz
Recommended
Keysight Model
909A Option 012
15