Second Harmonic Distortion (SHD)

Test Limits
This test checks the second harmonic distortion of the signal analyzer by
tuning to twice the input frequency and examining the level of the distortion
product. A low pass filter is inserted between the source and the signal
analyzer to prevent the source second harmonic from artificially raising the
second harmonic product displayed on the analyzer.
Item
Adapter, Type-N (m) to
BNC (f)
Adapter, BNC (f) to SMA
(m)
Adapter, 3.5 mm (f) to 3.5
a
mm (f)
Adapter, 2.4 mm (f) to 3.5
b
mm (f)
Cable, BNC 120 cm
(2 required)
Filter, 50 MHz Low Pass
Synthesized Sweeper
a. for Options 503, 508, 513, 526
Applied Frequency
40 MHz
a. Mixer Level = Input Level - RF Attenuation
Critical Specifications
Frequency: 10 MHz to 80 MHz
VSWR: < 1.05:1
Frequency: 40 to 80 MHz
Frequency: 10 MHz to 80 MHz
VSWR: < 1.1:1
Frequency: 10 MHz to 80 MHz
VSWR: < 1.1:1
Frequency: 10 to 40 MHz
Cutoff Frequency: 50 MHz
Rejection at 65 MHz: > 40 dB
Rejection at 75 MHz: > 60 dB
Frequency: 50 MHz
Spectral Purity: Better than –30 dBc
a
Distortion
Mixer Level
–15 dBm
< –60 dBc
(for this test)
19
Recommended
Keysight Model
1250-1476
1250-1200
83059B
11901B
10503A
0955-0306
PSG