Frequency Response (Flatness), Preamp On - Keysight Technologies X Series Manual

Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Keysight X-Series Signal Analyzers
PXA N9030A
Functional Tests
7

Frequency Response (Flatness), Preamp On

Test Limits
100 kHz to 3.6 GHz
> 3.6 GHz to 8.4 GHz
> 8.4 GHz to 13.6 GHz
> 13.6 GHz to 26.5 GHz
> 26.5 GHz to 34.5 GHz
> 34.5 GHz to 50 GHz
The frequency response test, with preamplifier on, measures the signal
analyzer's amplitude error as a function of the tuned frequency. Measurements
are made ranging from 50 MHz to the maximum frequency range of the
preamp. The signal source amplitude is measured with a power meter to
eliminate error due to source flatness. The measured value is normalized to 50
MHz.
Item
Adapter, Type N (m) to 3.5 mm
a
(m)
Adapter, 3.5 mm (f) to 3.5 mm
a
(f)
Cable, 3.5 mm, 1 meter
Frequency Range
Critical Specifications
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
Frequency: 10 MHz to 26.5 GHz
a
VSWR: < 1.4:1
Limit Relative to 50 MHz
±2.0 dΒ
±3.0 dΒ
±3.5 dΒ
±4.0 dΒ
±4.5 dΒ
±6.0 dΒ
(for this test)
35
Recommended
Keysight Model
1250-1743
83059B
11500E

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pxa n9030a

Table of Contents