Dräger Oxylog 2000 Instructions For Use Manual page 9

Emergency ventilator
Hide thumbs Also See for Oxylog 2000:
Table of Contents

Advertisement

If airway pressures are too high and there is a »Paw
high« alarm:
Check position of tube.
Check patient airways and use suction if necessary.
Ensure that the ventilation hose is not kinked.
To reset the warning message:
g
1 Press »
Reset« key.
2 Watch the pressure gauge so that faults in
ventilation can be detected quickly and danger to
the patient averted.
Applying PEEP
3 Set »PEEP« knob to the required value and check on
pressure gauge.
To override the mechanical stop for PEEP values greater
than 10 mbar:
Press and hold metal pin in scale and turn dial on
knob over pin at the same time.
Use the reverse proceedure to deliver PEEP values
below 10 mbar.
For cardio-pulmonary resuscitation
When applying cardiac massage for adults with the aid of
an assistant:
ª
4 Set »Freq.« knob to
ª
5 Set »Pmax« knob to
80 mbar.
Display:
Pmax= 80 mbar
CPR
g
6 Press »
Reset« to clear display.
The airway pressure is limited to max. 80 mbar without
interrupting inspiration prematurely (ventilation with
limited pressure).
A single beep is produced to advise that pressure
limitation is active.
= 12 1/min.
= pressure limited to
Using controlled ventilation IPPV
D
2
D
3
D
ª
4
ª
5