Perform device check

A
B
C
Connect the test lung
1 Connect the angled adapter (A) to
the breathing valve.
2 Connect the catheter connector
(B), diameter 7 mm, to the angled
adapter. The catheter connector
simulates the resistance of the
airways.
3 Connect the test lung (C).
Operation /Settings
Switch ON
1 To switch ON briefly press
the
key (B).
The device performs a self-test and
the operator is prompted, on the
display, to activate the configuration
menu or device check:
Oxylog 2000
plus
A
B
2 Press the rotary knob (A) to
confirm, before the bar is full.
3 Select Device check in the main
menu and confirm.
NOTE:
The device check can be
discontinued at any time by
pressing the »Alarm Reset« key.
26