Dräger Oxylog 2000 Instructions For Use Manual page 46

Emergency ventilator
Hide thumbs Also See for Oxylog 2000:
Table of Contents

Advertisement

Technical Data
ypical pneumatic operation time
at a minute volume of 10 L/min
11 L O
cylinder
2
2,5 L O
cylinder
2
Input voltage for Oxylog 2000
Connection for
external power supply
12 V / 24 V / 28 V DC
Fuse
(behind the cover of the battery
compartment)
Current consumption for high-speed
charging of the NiCd battery pack
Ventilator OFF
Ventilator ON
Current for trickle charging
Permissible ambient temperature
during charging
Electrical operation time
with internal NiCd battery pack
Alkaline manganese batteries
Protection class
Type of protection
Protection class of
power supply unit
Operating noise
Sound pressure level
Dimensions
(W x H x D) mm
Weight
O
cylinder, 2.5 L, full
2
O
cylinder, 2.0 L, full
2
Pressure reducer Alduk 1
NiCd battery pack
DC/DC converter
Operating environment
Electrical safety
Input voltage
Output voltage
46
Approx. 200 minutes without mixing (No Air Mix)
Approx. 400 minutes with mixing (Air Mix)
Approx. 45 minutes without mixing (No Air Mix)
Approx. 90 minutes with mixing (Air Mix)
12 V ±1 V DC
with DC/DC converter
F 1L 250V IEC 127 or
T 1L 250V IEC 127
300 mA (for 8 hours, then switches
automatically to trickle charging)
530 mA
30 mA
0 to 35 °C
Max. 6 hours at 5 to 50 °C
Max. 3 hours at temperatures below 5 °C
Max. 4 hours at 5 to 50 °C
Max. 2 hours at temperatures below 5 °C
with typical settings
m
Type BF
(body floating)
IP 54 (splash-proof)
in accordance with EN 60601-1
II
48 dB(A) at a distance of 1 m
215 x 123 x 208 (without handle)
4.3 kg
4.2 kg
3.5 kg
0.9 kg
0.15 kg
To be used only according to installation class 0, EN 61000-4-5
All voltages are in the SELV range, EN 60601-1
10.5 V DC to 30.0 V DC
12.5 V DC (+0.5 V / –1.0 V)