13.04 Self Check Error Screen
This screen shows a 3M™ Clean-Trace™ NGi Luminometer that has failed its Background
or Calibration check. Do not use the instrument to make measurements as performance
may be compromised. Contact 3M or local 3M distributor for service or calibration.
13.05 Other Error Screens
There is no available memory within the instrument to store further results. Press OK, upload
all stored results to the PC and try again.
There is an internal fault with the chamber or connections to the chamber. Contact 3M or
local 3M distributor.
Error Message
Unexpected error:
requesting results
Unexpected error:
uploading results
Not all sample plans have
been downloaded
Error when deleting and
downloading sample plans
Not all sample plans have
been downloaded.
Retry Upload Sequence
There are no Sample Plans
Available
Error received:
deleting results, results
have not been deleted
No Network Connection
13.06 Inadvertent Changing of Language Setting
In the event of the language setting being inadvertently changed, it may be difficult to return to the original language. If this occurs, follow the
following sequence of key presses to return to the original language:
1. Switch the 3M™ Clean-Trace™ NGi Luminometer off, then switch on and wait for main menu to appear.
2. Press the "up" navigation button once.
3. Press Right function button once.
4. Press the "down" navigation button once.
5. Press Right function button once.
6. Select desired language from list using navigation buttons and press Right function button once.
User Action
Disconnect the 3M™ Clean-Trace™ NGi luminometer from the USB lead, or remove from the docking
station, then reconnect and synchronise again.
Disconnect the 3M™ Clean-Trace™ NGi luminometer from the USB lead, or remove from the docking
station, then reconnect and synchronise again.
Using the 3M™ Clean-Trace™ Online Software administration website, check that sample plans have
been assigned to the 3M™ Clean-Trace™ NGi luminometer user.
Disconnect the 3M™ Clean-Trace™ NGi luminometer from the USB lead, or remove from the docking
station, then reconnect and synchronise again.
Disconnect the 3M™ Clean-Trace™ NGi luminometer from the USB lead, or remove from the docking
station, then reconnect and synchronise again.
Using the 3M™ Clean-Trace™ Online Software administration website, check that sample plans have
been assigned to the 3M™ Clean-Trace™ NGi luminometer user.
Disconnect the 3M™ Clean-Trace™ NGi luminometer from the USB lead, or remove from the docking
station, then reconnect and synchronise again.
Disconnect the 3M™ Clean-Trace™ NGi luminometer from the USB lead, or remove from the docking
station, then reconnect and synchronise again.
Indicates a network connectivity issue between the 3M™ Clean-Trace™ Online Software driver software
and the internet. The normal cause of this error is a proxy server installed on your network. Please contact
your IT helpdesk for further information.
SELF CHECK
High Background:
1,047 RLU
Calibration Check Failed.
Please leave the instrument at
stable temperature for 15 mins
and try again
Wed 04 Jan 2012 10:09
OK
Chamber Fault!
Unable to measure
Wed 04 Jan 2012 10:09
OK
OK
OK