1.0 PRODUCT DESCRIPTION
1.01 How the 3M™ Clean‑Trace™ NGi Luminometer works to Measure Contamination
1.02 ATP and Surface Hygiene Testing
1.03 ATP and Water Testing
1.01 How the 3M™ Clean‑Trace™ NGi Luminometer works to Measure Contamination
The 3M™ Clean-Trace™ NGi Luminometer is a luminometer used in conjunction with 3M reagent kits to measure levels of contamination
on surfaces and in water samples. The technology used is Adenosine Triphosphate (ATP) bioluminescence. ATP is a substance present in all
animal and vegetable matter, including most food debris, bacteria, fungi and other micro-organisms. Measurement of ATP is performed using
an enzyme reaction which occurs naturally in the tail of fireflies.
Luciferin/Luciferase (firefly reagent) + ATP = Light
The light emitted is in proportion to the amount of ATP. The intensity of the light emitted from a sample is measured by the
3M™ Clean-Trace™ NGi Luminometer and is displayed in Relative Light Units (RLU).
1.02 ATP and Surface Hygiene Testing.
The 3M™ Clean-Trace™ NGi Luminometer, together with the appropriate test kit, is an effective method for monitoring the hygiene status of
surfaces. ATP from both micro-organisms and product residues are detected. Measurement of "total ATP" provides an important insight to the
overall cleanliness. Product residues remaining on surfaces after cleaning will be a source of nutrients for any remaining micro-organisms
and may also protect the micro-organisms from disinfectant action. Because of the fast results, use of the 3M™ Clean-Trace™ NGi
Luminometer allows immediate remedial action to be taken if an unacceptable result is obtained. Surfaces can be re-cleaned before use,
ensuring good sanitation and contamination control.
The test kit instructions should be read carefully before use.
1.03 ATP and Water Testing.
Test kits are available for the assessment of contamination in liquid samples. Measurement of the total ATP in final rinse water samples is an
indicator of the effectiveness of Clean in Place (CIP) processes. The total ATP in process and waste water, water tanks and cooling towers can
be used to determine and monitor the efficiency of the biocide treatment program.
Measurement of free or non microbial ATP can be performed in addition to total ATP for further information on the nature of
the contamination.
The test kit instructions should be read carefully before use.

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Clean-Trace NGi Luminometer and is the answer not in the manual?

Table of Contents

Save PDF