Advertisement

Clean-Trace
Luminometer LX25
User Manual
IMPORTANT SAFETY INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
PRODUCT DESCRIPTION AND INTENDED USE: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
LIMITED WARRANTY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
OPERATING THE CLEAN-TRACE LUMINOMETER LX25 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
USER RESPONSIBILITY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
CLEAN-TRACE LUMINOMETER LX25 DESIGN AND COMPONENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
BATTERY USAGE AND MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
HOW TO PERFORM A QUICK TEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
COMMUNICATING WITH THE 3M QUALITY CONTROL DATA MANGER (QCDM) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
REGISTER A CLEAN-TRACE LUMINOMETER LX25 TO AN ORGANIZATION ON QCDM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
ROUTINE SYNCHRONIZATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
DE-REGISTRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
WI-FI SET UP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
VERIFYING DATA UPLOAD TO QCDM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
APPLICATION WORKFLOW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
STARTING THE AUDIT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
HOME SCREEN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
1 . FLEXIBLE ENDOSCOPE WORKFLOW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2 . ENVIRONMENTAL SURFACES WORKFLOW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3 . SURGICAL INSTRUMENT WORKFLOW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
PERFORMING A RETEST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
ADMINISTRATIVE FUNCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
ACCESSING ADMINISTRATIVE FUNCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
SETTINGS MENU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
CLEANING AND DECONTAMINATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
CONTACT US . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
CONTENTS
1
1
English
Issue Date: 2017-04

Advertisement

Table of Contents
loading

Summary of Contents for 3M Clean-Trace Luminometer LX25

  • Page 1: Table Of Contents

    COMMUNICATING WITH THE 3M QUALITY CONTROL DATA MANGER (QCDM) . . . . . . . . . . . . . . . .
  • Page 2: Important Safety Information

    Outside of the US contact your local 3M office . other manufacturers’ designs, or use in any other application that has not been evaluated or documented by 3M and may lead WEEE mark - This product contains electrical to an unsafe condition . The Clean-Trace Luminometer LX25...
  • Page 3 “Intrinsically Safe . ” User Manual after each use . − Do not use or place the Clean-Trace − Do not use the 3M Clean-Trace Luminometer LX25 with Luminometer LX25 on wet surfaces . ATP tests manufactured by companies other than 3M . • To reduce the risks associated with hazardous voltage −...
  • Page 4 − Do not use cleaning agents in the Clean-Trace Clean-Trace Luminometer LX25 in this User Manual . Luminometer LX25 chamber . − Use only 3M Clean-Trace ATP Tests and 3M accessories − Ensure the ATP Test is inserted into the Test sleeve and supplied by your local 3M sales representative activated before inserting the Test into the Clean-Trace or distributor .
  • Page 5: Limited Warranty

    REGULATORY COMPLIANCE LIMITED WARRANTY AND DISCLAIMER The CE Mark denotes compliance of the 3M™ Clean‑Trace™ 3M warrants that for one year from the date of sale, that Luminometer LX25 with the following European Directives: the Hardware will substantially perform in accordance with • R&TTE Directive 1999/5/EC – Radio and the User Documentation .
  • Page 6: Clean-Trace Luminometer Lx25 Design And Components

    1 . Power On/Off button shape is distorted . Battery replacement is performed by 2 . Cursor control buttons 3M Health Care Service Center or authorized 3M Service 3 . Touch screen Provider . See Care and Maintenance . 4 . USB/Power port Power On/Off/Sleep Mode...
  • Page 7: Inserting A Clean-Trace Atp Test Uxc Or H2O Into The Clean-Trace Luminometer Lx25

    • After the reading is complete, press the rocker cap to open the 3M™ Synchronization Manager Software installed on a the test chamber and remove the ATP Test UXC or H2O .
  • Page 8: Routine Synchronization

    5 . On the workstation, log into Synchronization 3 . When the landing page appears- Manager Software . STOP – remain on landing page – DO NOT LOG IN. 6 . Select the “Register” button on the workstation Synchronization Manager Software . 4 .
  • Page 9: Wi-Fi Set Up

    Before access to the workflow screens is available, an Synchronization Manager Software . organization hierarchy must be set up in QCDM, and the Set up Wi‑Fi Clean-Trace Luminometer LX25 must be registered to a QCDM organization . • Tap the “hamburger menu” in the upper right-hand corner of the Workflows are application specific .
  • Page 10: Home Screen

    “Recent failed audits” Tap “Perform Audit” • Select test point The 3M™Clean-Trace™ Luminometer LX25 has a unique workflow for each application . Please choose the appropriate application and follow the screen prompts . All available application workflow are outlined below .
  • Page 11: Environmental Surfaces Workflow

    2. Environmental Surfaces Workflow • Choose test point • From Home screen tap “Start” to begin new test or select a recent failed audit for retest • Insert ATP Test UXC or H2O • Select Area • Press “Begin Measurement” and wait until blue “Reading Result”...
  • Page 12: Surgical Instrument Workflow

    3. Surgical Instrument Workflow • Insert ATP Test UXC or H2O • From Home screen select “Start” • Press “Begin Measurement” and wait until blue “Reading Result” screen disappears then press “Return” • From select “Test Plan” screen, select “Routine Audit” • Repeat process until all test points have been tested .
  • Page 13: Administrative Functions

    Administrative Functions Settings menu Accessing administrative functions • Settings Menu Functions (detail) To access administrative functions, tap the “hamburger” menu at Wi-Fi Settings – turn Wi-Fi on/ the top, right-hand corner of the Home screen . off, set up network • Settings allow access to administrative functions Server settings –...
  • Page 14: Troubleshooting

    Use the Clean-Trace Luminometer LX25 at ambient temperature within normal operating range >40°C or <5°C between 5°C and 40°C . If the problem persists, contact 3M or your local distributor . Swab Left in Chamber The Clean-Trace Luminometer LX25 has a sensor to determine if a Clean-Trace ATP Test UXC or H2O has been left in the chamber for more than 30 seconds following the completion of a measurement .
  • Page 15: Cleaning And Decontamination

    Clean-Trace Helpline at 1-800-228-3957 . If any liquid is spilled inside the ATP Test chamber, do NOT attempt to clean inside the test chamber itself . Contact your 3M Health Care Service Center or local representative immediately . Clean and disinfect the external surfaces of the Clean-Trace Luminometer LX25 once per day at a minimum .
  • Page 16 Health Care Business St . Paul, MN 55144 USA Carl-Schurz-Str . 1 www .3M .com/Medical 41453 Neuss Germany 3M and Clean-Trace are trademarks of 3M . Used under license in Canada . © 2017, 3M . All rights reserved . Issue Date: 2017-04 34-8720-2654-6...