1.2 Front view
Alarms
Logo
Button 'UP'
Button 'LE'
'Fast access'
(see section 1.11.5)
1.3 Rear view
Option 1
1 2 3
Signal
(see section 1.5 and 1.6)
Detail of the plug-in screw terminals provided with
the instrument. The instrument is provided with all
terminals needed, both male and female.
1.4 Power connections
Earth connection - Although a
terminal is provided for earth
connection, this connection is
optional. The instrument does
not need earth connection for
correct operation nor for com-
pliance with the security regu-
lations.
Fuse - To comply with security regulation 61010-1, add to the power
line a protection fuse acting as disconnection element, easily acces-
sible to the operator and identified as a protection device.
Power 'H'
fuse 250 mA time lag
Power 'L'
fuse 400 mA time lag
Units
Button 'SQ'
'Configuration menu'
(see section 1.11)
Option 2
8 9 0
Power
(see section 1.4)
8 9 0
+
~
-
~
FEMA ELECTRÓNICA . Series S . S40-T
1.5 Signal connections - Pt100 / RTD
For Pt100 / RTD with 2 wires
1 2 3
Pt100 2 wires
For Pt100 / RTD with 3 wires
1 2 3
Pt100 3 wires
1.6 Signal connections - Thermocouple
1 2 3
1.7 Mechanical dimensions (mm)
36
72
+
-
18
80
Panel
32,5
cut-out
69
8
3