JDS Uniphase ORL-55 Series Operating Manual page 69

Optical return loss set
Table of Contents

Advertisement

Command
string
:MEM
:STOR
:MEAS
:MEM
:REC
:MEAS?
:MEM
:REC
:ALL?
:MEM
:DEL
:MEAS
:MEM
:DEL
:GROU
:MEM
:DEL
:ALL
JDSU
Parameter type / Response type /
Unit / Info
Stores the current measurement
under the current measurement
group and number.
Note: If this command is repeated
immediately (i.e. with no other
commands, such as
:MEM:REC:MEAS? in between), the
identifier increments by 1 for the next
store operation.
Returns the measurement stored in
the memory at the current group/
meas value.
Format for measurement:
group, meas; date, time;
measdata[; measdata]
Format of measdata:
wavelength, level, reference
e.g. 1,6;2007/AUG/31;23:59:59;
1490nm,-45.69dB,-5.12dBm;
1550nm,-12.34dB,-10.00dBm
Returns all the measurements stored
in the memory.
Format for each measurement:
see :MEM:REC:MEAS?
Deletes the current measurement.
Deletes all measurements in the
specified group (1...350).
e.g. :MEM:DEL:GROU 20
Deletes all measurement results from
the memory.
ORL-55
8 R
EMOTE
C
ONTROL
65

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orl-55/21Orl-55/22Orl-55/23

Table of Contents