JDS Uniphase ORL-55 Series Operating Manual page 67

Optical return loss set
Table of Contents

Advertisement

Remote command description
This command is only allowed in the RETURN LOSS
MODE with LASER ON. If the unit is in a different mode,
the unit will display an error message.
This command will switch the unit to user-specific
reference values.
NOTICE: This will take a couple of seconds.
If the unit is set to user-specific reference values, you will
see in the top line of hte display
Command
string
:RET:LOSS
:ENA:REF
:RET:LOSS
:ENA:REF?
JDSU
Return Loss C
Parameter type / Response type /
Unit / Info
<BOOLEAN>
Parameter type: BOOLEAN
0 = Switch to production reference
values (JDSU calibration)
1 = Switch to user-specific reference
values
(The command :EXEC:REF should
be executed prior to this command)
Parameter: none
Return value: BOOLEAN
0 = Production refrence values(JDSU
calibration) are set
1 = User-specific reference values
are set
ORL-55
8 R
EMOTE
C
ONTROL
63

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orl-55/21Orl-55/22Orl-55/23

Table of Contents