Remote command description
This command is only allowed in the RETURN LOSS
MODE with LASER ON. If the unit is in a different mode,
the unit will display an error message.
This command will switch the unit to user-specific
reference values.
NOTICE: This will take a couple of seconds.
If the unit is set to user-specific reference values, you will
see in the top line of hte display
Command
string
:RET:LOSS
:ENA:REF
:RET:LOSS
:ENA:REF?
JDSU
Return Loss C
Parameter type / Response type /
Unit / Info
<BOOLEAN>
Parameter type: BOOLEAN
0 = Switch to production reference
values (JDSU calibration)
1 = Switch to user-specific reference
values
(The command :EXEC:REF should
be executed prior to this command)
Parameter: none
Return value: BOOLEAN
0 = Production refrence values(JDSU
calibration) are set
1 = User-specific reference values
are set
ORL-55
8 R
EMOTE
C
ONTROL
63