JDS Uniphase ORL-55 Series Operating Manual page 41

Optical return loss set
Table of Contents

Advertisement

ORL MAIN menu
 Press [MENU ENTER] to open the ORL MAIN menu.
Normalize
Open Port
Manage
Memory
Configuration
Normalizing
To ensure a high measurement accuracy, it is required to
normalize the unit from time to time. This is done with the
port open, i.e. no DUT is connected.
To normalize the unit:
1. Disconnect DUT from test port.
The normalization can be performed only with an open
test port.
2. Unmount test adapter from test port, clean fibre
surface of test port and mount test adapter.
3. Switch on laser.
4. Press [MENU ENTER] to open the ORL MAIN menu.
JDSU
Stores the internal noise in ORL
mode as reference value.
Display and delete stored data (see
„Memory Management" on page
Change basic settings (see
overview" on page
Dangerous laser radiation!
Laser radiation can cause irreparable
damage to the eye and the skin.
!
When the system or test generator is
switched on, never look directly into the
output or into a connected optical fiber.
ORL-55
5 O
PERATION
„Device
9).
39).
37