JDS Uniphase ORL-55 Series Operating Manual page 50

Optical return loss set
Table of Contents

Advertisement

6 M
EMORY
Copying the result memory to a
memory stick
The result memory can be copied to a memory stick
plugged into the USB port on the connector panel. This
allows you to store a large amount of data in ASCII
format. The OFS-355 Download Manager can then be
used to transfer the result memory to a PC for further
evaluation or print out. For further details see
Download Manager" on page
Note: With OLT-55 and ORL-55, reading out the result
memory is not possible when running the units in laser
source mode.
To copy the result memory to a memory stick:
1. Plug in the memory stick to the device USB port on the
connector panel.
The folder structure is read.
Note: Depending on the number of folders and the
USB speed this procedure may take up to 30 seconds.
When completed the following message appears:
2. Press [MENU ENTER] to transfer the contents of the
measurement memory to the memory stick.
The contents are stored as a single file in a folder
named after the unit in use.
The measurement memory can now be deleted ready to
store new results.
Note: Do not remove the memory stick during the data
transfer.
A new data transfer can be started at any time as long as
the memory stick is plugged in.
46
M
ANAGEMENT
ORL-55
66.
„OFS-355
JDSU