Direct Mounting Of The Specimen Clamp On The Specimen Holder Fixture - Leica RM2125 RTS Instructions For Use Manual

Rotary microtome
Hide thumbs Also See for RM2125 RTS:
Table of Contents

Advertisement

4
Instrument Setup
To do so, proceed as follows:
1. Move the specimen holder fixture
handwheel
2. To release the clamp, turn the Allen screw
3. Push the guide
receptacle
4. To clamp the specimen clamp
as it will go.
4.5

Direct mounting of the specimen clamp on the specimen holder fixture

Note
The specimen clamps (standard or universal cassette clamps) can also be fastened directly to the
specimen holder fixture.
To do so, proceed as follows:
1. Move the specimen holder fixture
handwheel and engage the handwheel lock.
2
3
Fig. 9
2. Then remove the dovetail receptacle
To do so, loosen and unscrew the four screws
(→ Fig.
10-1).
1
4
Fig. 10
22
(→ Fig. 8-1)
(→ Fig. 8-6)
and engage the handwheel lock.
(→ Fig. 8-4)
of the specimen clamp
(→ Fig. 8-5)
as far as it will go.
(→ Fig. 8-3)
(→ Fig. 9-1)
1
2
3
to the upper end position by turning the
(→ Fig. 8-2)
counterclockwise.
(→ Fig. 8-3)
turn the Allen screw
to the upper end position by turning the
(→ Fig. 9-2)
from the specimen holder fixture
(→ Fig. 9-3)
using an Allen key No. 3
from the left into the dovetail
(→ Fig. 8-2)
clockwise as far
(→ Fig.
Version 2.6, Revision J
9).