4.3

Unpacking and installation

Note
The packaging has two indicators, ShockDot Impact Indicator and Tilt Indicator, that indicate
improper transport. When the instrument is delivered, check these first. If one of the indicators
is triggered, the package was not handled as prescribed. In this case, please mark the shipping
documents accordingly and check the shipment for damage.
Fig. 6
Warning
Do not transport the instrument by holding it by the handwheel grip, coarse feed wheel, or rotary
knob for setting the section thickness!
Leica RM2125 RTS
1
• Remove the packing strap and the adhesive
tape
(→ Fig.
6-1).
• Remove the carton lid
2
• Take out the accessory carton
• Take Allen key No. 8 out of the accessory
carton and put it aside for later use.
• Remove the fixing carton
• Remove the outer carton wall
• Use both carrying straps
front and rear to remove the instrument on the
wooden pallet from the carton.
3
4
5
6
Instrument Setup
(→ Fig.
6-2).
(→ Fig.
6-3).
(→ Fig.
6-4).
(→ Fig.
6-5).
(→ Fig. 6-6)
at the
4
19