Pressing the Esc key, you will go back to the Main Menu, with cursor ready to the next item.
By selecting the OUT MODE item and pressing the touching Enter key, the relative window appears.
Here following the meaning of each item of this menu:

Out Mode

11.2.4.
- NATIVE
The Transfer Function is transparent (no correction is applied)
- DICOM (FACTORY) The output curve follows the DICOM curve pre-setted in the factory
- DICOM (USER)
The output curve follows the DICOM curve pre-setted in the factory
and an offset can be applied to the curve (see DICOM OFFSET)
- COLOR (FACTORY) The monitor output is corrected according to the color temperature
pre-setted in the factory (6500°K).
In the menu, the string „COLOR‟, is replaced with the current color
temperature, i.e.
- COLOR (USER)
The monitor output is corrected according to the color temperature
selected by the user (5600, 6500, 7600, or 9300°K, see COLOR TEMP)
In the menu, the string „COLOR‟, is replaced with the current color
temperature, i.e.
Gamma correction is available in this mode, see GAMMA COMP
the selected mode will be loaded in 'real time' on the monitor, so you can see image modifications
NOTE:
moving the cursor.
MONITOR INFO
-> OUT MODE
USER SETTING
OSD SETTING
PRODUCT INFO
SERVICE
TRANSFER FUNCTION
-> NATIVE
DICOM (FACTORY)
DICOM (USER)
COLOR (FACTORY)
COLOR (USER)
____6500°K (FACTORY)____
____7600°K (USER)____