Error no..
10
Overtemperature
In the rotor chamber (more than + 50
Reason:
Breakdown of the
refrigeration system / Electronic faiiure
Action;
Call service
Reason:
Temperature sensor detective I Electronic faiiure
Actión:
Call service
Error no.: 11
Temperature sensor
Reason:
Short circuit at the temperature sensor or at the sensor cable
Action:
Caü service
Reason:
Chamber temperature is too low, below - 25°C
Action:
Solenoid valve is not werking. Call service
Error no.: 20
No rotor Identification
Reason:
No rotor inserted
Action:
Insert rotor into the unit
Reason:
Rotor Identification sensor detective
Action:
Call service
Reason:
Inserted rotor has no indicator ring
Action:
Use a correct rotor
Reason:
Rotor is not fixed correctly to the motor shaft
Action:
insert rofor correctly. The pins has to align correctly with the
rotor slots (see chapter 3)
Error no.: 21
Start Bit is missing
Reason:
A magnet of the indicator ring is missing
18 E
•C)
Z 300/Z 400/Nov. 1996
6 D
1.4
Technische Daten
Hersteller
Hermle Labortechnik GmbH
Typenbezeichnung
Z400
Abmessungen:
Breite
41 cm
Tiefe
37,5 cm
Höhe
56,5 cm
Gewicht
55 kg
Gerauschpegel max.
60 dB(A)
max. Drehzahl
13.500 min'
max. Volumen
1000 ml
max. RZB
17.310 xg
zulassige Dichte
1,2 kg/dm
zulassige kinetische Energie
4000 Nm
elektr. AnschluR
230 V/50-60 Hz 1ph
Stromaufnahme
3.8 A
Leistungsaufnahme
650 Watt
Funkentstörung
VDE 0871, Funkentstörgrad B
Prüfpflicht nach UW VBG 7z
ja
Vom Betreiber einzutragen:
Inventar-Nr.:
Überwachungs-Nr.:
Ort der Aufstellung:
Wartungsvertrag:
Service Adresse
Hermle Labortechnik
Gosheimer Str. 56
78564 Wehingen
Tel.:07426 / 96 22 55
Fax.:07426 / 96 22 49
zustandiger Handler:
3
120 V/50-60 Hziph
4,0 A
480 Watt
Z 300/2 400 Juli 1996