Internal Span Adjustment; External Span Adjustment; External Span Adjustment With Manual Input Of Weight Values - Sartorius Master Pro Series Service Manual

Electronic precision balances
Table of Contents

Advertisement

Max5200 g
0.00
g
int. adjust
c
int. adjust
Note:
0.09
g
int. adjust
Max5200 g
0.00
g
Note:
Max5200 g
202122
202122
S
info
2000.02
g
S
cal./adj. weight
14

Internal Span Adjustment

All LP balances are equipped with a built-in motorized calibration weight.
– Turn on the balance and allow it to warm up.
d= 0.01 g
– Zero the balance, if necessary (press t).
– Press q; press v to select "int. calibr." or "int. adjust" and press
q to start the internal span adjustment.
wahl
– The built-in calibration weight is applied automatically, the weight is
stored, and the deviation from the last span adjustment is displayed.
Depending on the menu code setting (under 1 10 X), either calibration/ad-
justment is made automatically (code 1 10 1: calibrate, then auto-adjust), or
the deviation from the last calibration is displayed, but no adjustment is
made until you press a key (code 1 10 2: calibrate, then manual adjust).
– If the error exceeds the tolerance limits, press q to make the correspond-
ing adjustment, or press c to cancel the adjustment procedure.
– The balance then returns to the SELECT mode.
end

External Span Adjustment

There are two ways to perform external span adjustment:
d= 0.01 g
1. with the Sartorius service software (PSION Server or SARTOCAS Service
Software for PCs) - if the adjustment error is greater than
2. with the balance in the service mode, as described in this manual.

External Span Adjustment with Manual Input of Weight Values

You only need to input the calibration weight for span adjustment if:
1. the available weight is not sufficiently accurate (at least one digit greater
than the display accuracy), or
2. you are working with a weight other than the recommended weight (e.g.,
with a heavier weight, close to the maximum capacity).
– Press e to turn on the balance.
d= 0.01 g
– The balance performs a self-test.
– When 0.00 g is displayed (depending on balance model), enter the ser-
S-ID
vice password "202122" and press M.
– Press v to activate the INPUT mode.
– Press Y (several times) to select the "cal./adj. weight" menu item.
menU input
– Enter the conventional weight value, including all decimal places, using
the numeric keys.
– Press M to store the adjustment and exit the SETUP menu.
±