Max5200 g
0000.00
g
S
Reprotest
Max5200 g
0000.00
g
S
disp. x 10
16
reproTEST
– Activate the service mode (see page 11).
– Press v (several times) to select the "reprotest" menu item.
d= 0.01 g
This tests the reproducibility of the automatic application of the built-in calibra-
selec.
tion weight.
The internal weight is applied, measured and removed 6 times.
At the end of the routine, the reproducibility value is displayed.
High Resolution - Display * 10
– Activate the service mode (see page 14).
– Press v (several times) to select the "disp x10" menu item.
d= 0.01 g
The balance switches to SuperRange resolution and an additional digit is dis-
selec.
played (for test purposes and initial verification).