• If the initialization process fails, check the following.
- Check drill fixing sponge removed
- Check carriage fixing sponge removed (Right and Left)
- Check carriage fixing bracket removed
- Check feeler fixing tape removed
• If the touch screen does not work, run the touch calibration function.
- Turn off the system and then turn on the system with the [WINDOW] key
pushed.
- After a while touch calibration screen displayed.
- Follow the instruction displayed on the screen.
• If the initialization error occurs continuously, follow the instruction below.
- Memorize error message displayed on the screen.
- Turn off the system and then turn on the system with the [STOP] key pushed.
- After a while main screen displayed on the screen without initialization.
- Run "Tool – Test Mode" in the menu bar and check the system.
• If the edger work correctly, make sure reset current time. (see chapter [7.1] for
setup time)
• Make sure to turn off the system before insert or remove SD memory card.
• If you insert or remove the SD memory card while edger is running, the data in
the SD card can be corrupted and that data corruption can cause machine
malfunction.