Installation

The HMIA is quickly and easily installed.
Prior to installation of the HMIA, verify that the console has the correct software to support the HMIA. The
software version is displayed during initial start-up of the console per Figure 4.75. The first number in the Soft-
ware Version must be equal to or greater than 1.04 (See Figure 4.75).
Figure 4.75. Console Software Version
If your console does not have the compatible software version, then the correct software will need to be
installed for the HMIA to function. Failure to install the correct software prevents the HMIA from functioning
but it has no adverse affects on the console.

Installation Equipment

The following items are necessary to install a HMIA.
1. Console with correct software.
2. Hospital Monitor Interface Accessory (HMIA)
3. TP-400 test fixture with adapter cable that mates to the T1 connector of the console
4. Air trap assembly filled with water.
5. The standard Hospital Monitor patient temperature cable that is used to connect the Hospital Monitor to a
Foley or rectal patient temperature sensor.
6. Hospital Monitor that conforms to the YSI-400 requirements.
7. Optionally a Digital Multimeter may be used if a hospital monitor is not available.

Installation Procedure

1. Disconnect power to the console.
2. Disconnect the Umbilical Cable from the rear of the display head (See Figure 4.76).
78