Multi-Function Electrodes
INTRODUCTION
ZOLL PD 2200 Multi-function electrodes allow the operator to "hands-off"
defibrillate, cardiovert, pace, and ECG monitor with a single pair of electrodes.
The PD 1400 has been designed to operate with standard paddles and ZOLL NTP
2000 pacing (standard) electrodes or PD 2200 multi-function electrodes without
standard paddles. To switch between standard paddle-electrode/multi-function
electrode options, the operator must remove either the standard paddles-pacing
cable or the multi-function cable and install the other. (See photos on page 12.)
MULTI-FUNCTION
ZOLL PD 2200 Multi-Function Pacing/Defibrillation Electrodes are anterior1
ELECTRODES
posterior, pre-gelled, disposable electrodes. They are applied to the patient in the
same manner and location as standard pacing electrodes. The multi-function
electrodes can only be used with a multi-function cable. (ZOLL NTP 2000 pacing
electrodes can only be used with a standard pacing cable.)
WARNING
The ZOLL PD 1400, cables, and electrodes are designed and tested as a unit to
provide maximum patient safety and comfort. Use of electrodes and/or cables
other than those supplied by ZOLL could compromise patient safety
If continuous pacing with multi-function electrodes exceeds 8 hours, you
must replace the electrodes.
MULTI-FUNCTION
The ZOLL PD 441 0 Multi-Function Cable connects
CABLE
to the standard paddles receptacle on the right
front of the unit in the same manner as the
paddles cable. To remove either cable, press the
two locking buttons -one on top of the connector
and one on the bottom -together and pull the
connector out from the PD 1400. To install either
cable, simply plug it into the connector.
When using the multi-function cable, the operator must check the electrode
packaging labels. Only "Pacing/Defibrillation Electrodes" will operate with the
multi-function cable. The labels on the electrodes must have the words "PACE"
printed in green and "DEFIB" printed in red.