ZOLL IVTM Operation Manual page 95

Intravascular temperature management
Hide thumbs Also See for IVTM:
Table of Contents

Advertisement

TEMPERATURE PROBES DO
1
NOT AGREE
SYSTEM MALFUNCTION
Table 5.2. Medium priority Alarm Messages, Causes, and Corrective Actions (continued)
1. Stops operation of the console.
Alarm Message
DOWNLOAD ERROR. PLEASE
CHECK EXTERNAL
COMPUTER
PATIENT DATA LOG FULL.
PLEASE DOWNLOAD NOW
Table 5.3. Low priority Alarm Messages, Causes, and Corrective Actions
Temperature outputs from the
primary and secondary patient
temperature probes differ by
more than 2°C.
May be caused by a variety of
1
failures.
Cause
Patient trend data download
has failed or there is no com-
puter connected to the con-
sole.
Patient trend data log is full.
Only two hours of space
remain.
Determine the cause of the discrepancy
and correct it. One or both temperature
probes may need to be replaced.
Discontinue use and contact your ZOLL
service representative.
Corrective Action
Check the connection between the
computer and the console and select
"Try again." If problem persists, select
"Cancel" and contact ZOLL technical
support.
Download patient trend data.
If data is not downloaded, it will be
overwritten.
87