Temperature Trend Data; Overview; Displaying The Temperature Trend Graph - ZOLL IVTM Operation Manual

Intravascular temperature management
Hide thumbs Also See for IVTM:
Table of Contents

Advertisement

7. Lift the handle on the pump rollers.
8. Grasp the pump tubing and gently pull it up and out of the channel while rotating the pump head.
9. Pull the tubing out of the pump head. Press the handle down onto the rollers until it press into the detent
and close the top cover of the roller pump.
10. Loosely coil the disconnected ends of the catheter tubing.
11. Lift the heat exchange coil out of the coolant well. Hold the coil above the coolant well to allow coolant to
drip back into the well.
12. Pull the air trap up and out of the holder.
13. Unhook the saline solution container from the hook.
14. Bundle up the collection of components, still connected together, and gently deposit them in the medical
waste container. Take care to avoid contact with the IV container "spike."
15. Use a disposable tissue to wipe up any spilled coolant from the top of the coolant well. Place the tissue in
the medical waste container.
16. Replace the insulated cap on the coolant well. Replace the insulating jacket on the hook.
Removal and disposal of the used components is complete. The console may now be stored or moved to its
next treatment location.
Note. Disposal of console. When the console itself has reached the end of its useful life, it must be disposed
of in accordance with local governing ordinances and recycling plans for refrigerated appliances.

Temperature Trend Data

Overview

During operation, the console continuously collects and stores temperature trend data, storing a record each
minute. This data is stored in memory and can be downloaded to an attached laptop computer for later analy-
sis or plotting.
Data can be collected for 21 days before the memory is filled. Periodic data download and deletion is recom-
mended to preserve all of the stored data. If data is not downloaded before the memory is full, an error mes-
sage (M Id:16) appears and the console stops treatment. The data should immediately be deleted in order to
continue.
Note. M Id:16 Error Message. If during treatment the M Id:16 error message appears, this indicates the
treatment file memory is full. The console stops treatment. To continue treatment, power the console off and
on. Follow the prompts and select current settings. When the data download information appears, immedi-
ately either delete the data or use TempTrend to download the data and delete it. You may then resume treat-
ment.
At any time during operation, the complete record of temperature trend data can be displayed as a graph on
the screen. This chapter explains how to display temperature trend data and provides details about the format
and structure of the downloaded data.

Displaying the Temperature Trend Graph

To display the temperature trend graph, follow these steps:
1. Press the Press for Menu/Enter knob once. The screen displays the menu. The selection "View Graphs" will
be highlighted.
69

Advertisement

Table of Contents
loading

Table of Contents