Fever (FVR)
In this treatment option, the console will start cooling the patient once the patient temperature is above the
target temperature. It does this by keeping the bath at its coldest permissible temperature and then operating
the peristaltic pump whenever the patient's temperature moves above the target temperature.
WARNING. "Lo" patient temperature alarm limit with "Fever". The console will not warm the patient
when the "Fever" treatment option has been selected. The "Lo" patient temperature alarm limit ensures that
an alarm occurs should the patient stop regulating his/her own body temperature. Such patients will cool to
room temperature. This can occur when the patient dies or becomes comatose. Investigate all patient tem-
perature alarms.

User Interface

The primary controls are mounted in the display head of the console. The next sections describe:
Controls and display screen
Menu system
Console Controls
The controls and display screen are illustrated in Figure 2.2 on page 13. The use of each control is described in
the table below.
Power Indicator
Alarm Indicator
Standby/Run button
Target Temperature button
Rate Deg/Hr button
Press for Menu/Enter knob
Table 4.2. Console Controls

Display Screen

The console features a color display screen. Figure 4.3 shows the screen as it appears during Run.
28
Glows green when the display screen has power.
Illuminates red during error conditions that prevent the con-
sole from displaying an alarm message.
Toggles between Run and Standby
Press this to change the target temperature
Press this to change the rate and/or treatment mode.
This knob operates the menus on the display screen. Rotate
the knob to scroll between selections. Press and release the
knob to select.