Figure 4.42. Replace the Cap on the Coolant Well
Spike the Saline Bag
15. Using aseptic technique, connect the priming line to the sterile saline container. The priming line is
equipped with a "spike" connector. Hang the saline container on the hook provided.
Figure 4.43. Making a Sterile Connection to the Saline Container
Caution. Start-Up Kit spike. The spike on the Start-Up Kit is relatively long. Be careful not to puncture the
side wall of the saline bag when connecting to the Start-Up Kit.
Prime & Fill the Air Trap
16. Remove the air trap from its holder and hold it upside down (with the tubing connections pointing
downward).
55