Figure 7.1. Location of Condensate Pan (see arrow).
5. The pan may be washed in hot soapy water. Never use solvents or abrasive cleaners on the pan. Dry the
pan when finished and reinstall it in the console.
Note. Do not operate the console without a condensate collection pan properly installed. Failure to do so may
allow water to accumulate on the floor under the console.

Drain Coolant

It may become necessary to drain and clean the coolant well because the coolant has become contaminated,
or to prepare the console for shipment.
Required Tools and Materials
You will need the drain line and connector assembly (supplied with the console), and a 2-liter waste container
to perform this task. Drain the coolant in an area where spills can be managed appropriately.
Procedure
To drain the coolant well, follow these steps in the indicated order.
1. Unplug the console.
2. Remove the cap on the coolant well and set it aside on a clean surface.
3. Visually inspect the coolant well for discolorations or foreign matter.
4. Remove the front panel of the console by pulling the panel straight away from the front of the machine
(no tools are required to remove the panel). Near the bottom left corner of the console is a drain hose held
by a clamp and the drain coupler.