ZOLL IVTM Operation Manual page 12

Intravascular temperature management
Hide thumbs Also See for IVTM:
Table of Contents

Advertisement

Immunity Test
Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency (50/
60 Hz) magnetic field
IEC 61000-4-8
Table 1.2. Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Thermogard XP console is intended for use in the electromagnetic environment specified below. The
customer or user of the console should ensure that it is used in such an environment.
Immunity Test
Conducted RF
immunity
IEC 61000-4-6
4
IEC 60601 Test
Compliance Level
Level
Voltage Dips >
Voltage Dips >
95% reduction,
95% reduction,
0.5 period At 0°,
0.5 period At 0°,
45°, 90°, 135°,
45°, 90°, 135°,
180°, 225°, 270°
180°, 225°, 270°
and 315°
and 315°
Voltage Dips >
Voltage Dips >
95% reduction, 1
95% reduction, 1
period At 0°
period At 0°
Voltage
Voltage
Interruptions >
Interruptions >
95% reduction,
95% reduction,
250/300 periods
250/300 periods
30 A/m
30 A/m
IEC 60601 Test
Compliance Level
Level
3 Vrms
3 Vrms
150 kHz to
150 kHz to 80 MHz
80 MHz
6 Vrms in ISM
6 Vrms in ISM
bands within
bands within
150 kHz and
150 kHz and
80 MHz (see Note
80 MHz
3)
80% AM at 1 kHz
80% AM at 1 kHz
Electromagnetic environment --
guidance
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If you require continued operation
during power mains interruptions, we
recommend that the console be powered
by an uninterruptable power supply or a
battery.
Power frequency magnetic fields should
be at levels characteristic of a typical loca-
tion in a typical commercial or hospital
environment.
Electromagnetic environment
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the console, including cables,
than the recommended separation dis-
tance calculated from the equation appli-
cable to the frequency of the transmitter.
Recommended separation distance
150 kHz to 80 MHz
d = 1.2
, 80 to 800 MHz
d = 2.3
, 800 MHz to 2.7 GHz

Advertisement

Table of Contents
loading