Accidental Disconnection T1/T2 Probe

Disconnection of the T1 probe during Run results in an
alarm. Press the knob to silence the alarm. The console
moves to Standby. Treatment cannot continue until the
T1 probe has been replaced.
Note that the patient temperature is displayed as
"---"and the yellow warning banner covers the bath
temperature display.
Absence or disconnection of the T1 probe during
Standby results in a warning display without a per-
sistent audible alarm. The console cannot enter Run
until a T1 probe has been connected.
Note that the patient temperature is displayed as
"---"and the yellow warning banner does not cover the
bath temperature display.
Disconnection of the T2 probe during Run results in an
alarm. Press the knob to silence the alarm. The console
moves to Standby. If you attempt to return to Run
without reconnecting the T2 probe, you will be asked
to verify your intention.
Note that the patient temperature will be displayed
correctly if the T1 probe remains connected. The yellow
warning banner covers the bath temperature display.
Alarms & Alerts
A detailed description of the alarms of the console is provided in later sections of this manual. See:
Alarms and Corrective Actions on page 85
Troubleshooting on page 89

Alerts

Alerts can be cleared by correcting the problem that caused them to occur. For example, a low coolant alert
can be cleared by adding coolant to the coldwell of the console. In most cases, the console will identify and
notify you to rectify alert states at power up.