Preparing the Console for Treatment

To prepare the console for treatment, follow these steps in the indicated order.
1. Roll the console to a convenient position near the patient's bedside. Plug the power cord into a hospital-
grade receptacle.
2. Lock the right front caster by stepping down on the tab above the wheel.
3. At the rear of the console, near the upper left corner is the power on/off switch. Turn the switch ON.
Figure 4.21. Power Switch and Power Indicator Lamp
4. The green power indicator lamp will be illuminated and the alarm will beep one long beep followed by a
shorter softer beep.
5. The console performs a self-test. During this interval, the self-test screen appears on the display screen. If
the self-test detects a problem, an error message will be displayed. If this occurs, refer to Alarms and
Corrective Actions on page 85 for assistance. At the end of the self-test there will be two short beeps.
Note. The software version shown in Figure 4.22 is shown for reference only; the current product may use
a different software version.
Figure 4.22. Self-Test Screen
6. When the self-test is finished, the System Set Up screen displays the message "Bath Pre-Set?"
44