Operation and maintenance
May 2020
Note
Use the Rosemount Wireless LCD display part number: 00753-9004-0002.
56
Diagnostic Button Screen 8: the device cannot retrieve
information from the radio in the device - the device may
still be operational and publishing HART data
r a d i o
f a i l u r
Diagnostic Button Screen 9.1: configuration of the
transmitter is invalid such that critical operation of the
device may be affected - check the extended configuration
c o n f g
status to identify which configuration item(s) need to be
corrected
f a i l u r
Diagnostic Button Screen 9.2: configuration of the
transmitter is invalid such that non-critical operation of
the device may be affected - check the extended
c o n f g
configuration status to identify which configuration
item(s) need to be corrected
w a r n
Diagnostic Button Screen 10.1: a sensor attached to the
transmitter has failed, and valid readings from that sensor
are no longer possible - check the sensor and sensor wiring
s n s r
connections - check additional status for more detailed
information of the failure source
f a i l u r
Diagnostic Button Screen 10.2: a sensor attached to the
transmitter is degraded, readings from that sensor may
not be within accuracy specifications - check the process,
s n s r
and sensor wiring connections - check additional status for
more detailed information of the warning source
w a r n