Configuration
May 2020
code is specified. The "C1' option code requires user supplied information of process pipe
material and pipe schedule. Rosemount X-well technology can be configured with any
asset management software that supports Electronic Device Description Language
(EDDL). The Device Dashboard interface with DD revision 648 Dev. 4 Rev. 1 or higher is
required to view Rosemount X-well functionality.
The "Rosemount X-well Process" sensor/type option should be selected as the sensor type
in most cases. Once selected, pipe material, line size, and pipe schedule information is
required when configuring Rosemount X-well technology. This section is referring to the
process pipe properties that Rosemount 648 Wireless and 0085 Pipe Clamp Sensor with
Rosemount X-well technology is going to be installed in. This information is required for
the in-transmitter algorithm to accurately calculate process temperature.
In the rare case that the process pipe is not available, a custom value for the pipe
conduction coefficient can be entered. This field becomes available when the "Rosemount
X-well Custom" sensor/type option is selected.
Configure Rosemount X-well technology with Field
Communicator
Procedure
1. From the Home screen, select 2: Configure.
2. Select 1: Guided Setup.
3. Select 1: Configure Sensor.
4. Select 1: Configure Sensor Type and Units.
5. Select either Rosemount X-well Process or Rosemount X-well Custom.
6. Select desired configurations and select Enter.