Reference Manual
00809-0200-4648
Figure 3-1: Antenna Position
3.3.3
Conduit entry
Upon installation, ensure each conduit entry is either sealed with a conduit plug using
approved thread sealant, or has an installed conduit fitting or cable gland with appropriate
threaded sealant.
Note
The conduit entries on the Emerson 781 Field Link are threaded ½–14 NPT.
Figure 3-2: Conduit Entry
A. Conduit entry
3.3.4
Field Communicator connections
The Black Power Module needs to be installed in the device for the Field Communicator to
interface with the Rosemount 648 Wireless. For HART Wireless Transmitter
communication via a Field Communicator, a Rosemount 648 Wireless Device Dashboard
(DD) is required. Rosemount 648 Wireless Transmitters equipped with Rosemount X-well
Technology requires DD revision 648 Dev. 4 Rev. 1 or higher to view Rosemount X-well
functionality. To obtain the latest DD, visit the Field Communicator System Software and
Device Description site at:
Refer to
Rosemount 648 Wireless Transmitter.
Rosemount 648 Wireless
Emerson.com/Field-Communicator
Figure 3-3
for instructions on connecting the Field Communicator to the
Installation
May 2020
35