Figure 1-9 Offset Accuracy Verification Setup - Keysight 8990B Service Manual

Peak power analyzer
Hide thumbs Also See for 8990B:
Table of Contents

Advertisement

1
Performance Verification and Adjustments
Test procedure
1 Assemble the test equipment as shown in
Figure 1-9
2 Set the digital multimeter to measure DC voltage.
3 Press Default Setup on the 8990B to set the 8990B to its default settings, and
4 Enable averaging and set the average count to 256.
5 Set the input impedance for the video input channel 2 to 50 Ω.
6 Set the vertical scale for the video input channel 2 to 10 mV/div.
7 Set the video input channel 2 offset to 120 mV.
8 Set the DC power supply to output ±120 mV as shown in
9 Press Clear Display, and wait for the average count to reach 256.
10 Record the voltage reading at the digital multimeter in V
11 Record the mean value of the average reading at the 8990B in V