4
Repair Guide
Check Source Board Module Disassembly
Use this procedure to disassemble the check source board module.
1 Detach the U-wave attenuator pad (0955-0462) on the check source board
Figure 4-49
2 Unscrew the four screws (0515-0390) with the 18 in-lbs, T20 Torx screwdriver.
132
assembly using the manual torque wrench (5/16") with torque of 5 in-lbs as
shown in
Figure
4-49.
Detach the U-wave attenuator pad
Detach the belt cover with the poron pad (08990-00110) as shown in
Figure
4-50. Unscrew the four screws (0515-2007) with the
5 in-lbs, T8 Torx screwdriver and then remove the check source assembly
(08990-66004) from the PPMC assembly.
Keysight 8990B Service Guide