1
Performance Verification and Adjustments
Table 1-4
Test record for offset accuracy verification (continued)
Vertical scale
Output
(V/d iv)
vol tage
Video input channel 3 (50 Ω)
0.01
±120 mV
0.02
±240 mV
0.05
±600 mV
0.1
±1.2 V
0.2
±2.4 V
0.5
±4.0 V
1
±4.0 V
Video input channel 3 (1 MΩ)
0.01
±5 V
0.02
±10 V
0.05
±10 V
0.1
±20 V
0.2
±20 V
0.5
±20 V
1
±100 V
2
±100 V
5
±100 V
42
V
V
V
DMM+
scope+
error+
V
V
V
DMM–
scope–
error–
Keysight 8990B Service Guide
V
zero-error
V
zero-error
limit (±)
1.8 mV
2.6 mV
5.0 mV
9.0 mV
17.0 mV
41.0 mV
81.0 mV
1.8 mV
2.6 mV
5.0 mV
9.0 mV
17.0 mV
41.0 mV
81.0 mV
161.0 mV
401.0 mV

Advertisement