Reference Manual
May 2019
URV= 5833 SCFM
= 8.36 mA
Therefore, an input frequency of 200 Hz represents a current output of 8.36 mA.
Rosemount 8600 Vortex Flow Meter
LRV= 0 SCFM
F Hz
K C
I
=
20
--------------------------------------------------------------- -
URV LRV
–
200
10.797 0.011641
--------------------------------------------------------------------------------- -
I
=
5833 0
F
(Hz) = 200.00
in
–
LRV
16
+
4
0
–
–
Electronics Verification
00809-0100-4860, Rev BF
16
+
4
113