Pipe length requirements
The vortex meter may be installed with a minimum of ten
diameters (D) of straight pipe length upstream and five
diameters (D) of straight pipe length downstream.
Rated Accuracy is based on the number of pipe diameter
from an upstream disturbance. No K-factor correction is
required if the meter is installed with 35 D upstream and 10
D downstream.
Tagging
The flowmeter will be tagged at no charge, according to
customer requirements. All tags are stainless steel. The
standard tag is permanently attached to the flowmeter.
Character height is 1/16-in. (1,6 mm). A wired-on tag is
available on request. Wire on tags can contain five lines
with up to 28 characters per line.
Flow calibration information
Flowmeter calibration and configuration information is
provided with every flowmeter. For a certified copy of flow
calibration data, Option Q4 must be ordered in the model
number.