Emerson Rosemount 8600 Series Reference Manual page 102

Vortex flow meter
Hide thumbs Also See for Rosemount 8600 Series:
Table of Contents

Advertisement

Specifications and Reference Data
May 2019
Table A-11. Saturated Steam Flow Rate Limits
Process
Flow Rate
Pressure
Limits
15 psig
(1,03 bar G)
25 psig
(1,72 bar G)
50 psig
(3,45 bar G)
100 psig
(6,89 bar G)
150 psig
(10,3 bar G)
200 psig
(13,8 bar G)
300 psig
(20,7 bar G)
400 psig
(27,6 bar G)
500 psig
(34,5 bar G)
1.
Assumes steam quality is 100%
Table A-12. Saturated Steam Flow Rate Limits
Process
Flow Rate
Pressure
Limits
15 psig
(1,03 bar G)
25 psig
(1,72 bar G)
50 psig
(3,45 bar G)
100 psig
(6,89 bar G)
150 psig
(10,3 bar G)
200 psig
(13,8 bar G)
300 psig
(20,7 bar G)
400 psig
(27,6 bar G)
500 psig
(34,5 bar G)
1.
Assumes steam quality is 100%
94
(1)
Minimum and Maximum Saturated Steam Flow Rates
for line sizes 3-in./DN 80 through 4-in./DN 100
3-in./DN 80
Rosemount 8600D
lb/hr
max
3330
min
298
max
4370
min
341
max
6914
min
429
max
11874
min
562
max
16763
min
668
max
21630
min
759
max
31389
min
914
max
41258
min
1073
max
51297
min
1334
(1)
Minimum and Maximum Saturated Steam Flow Rates
for line sizes 6-in./DN 150 through 8-in./DN 200
6-in./DN 150
Rosemount 8600D
lb/hr
max
13013
min
1163
max
17080
min
1333
max
27019
min
1676
max
46405
min
2197
max
65611
min
2610
max
84530
min
2965
max
122666
min
3572
max
161236
min
4192
max
200468
min
5212
–Line Sizes 3- through 4-in.
4-in./DN 100
Rosemount 8600D
kg/hr
lb/hr
1510
5734
135
513
1982
7526
155
587
3136
11905
195
739
5386
20448
255
968
7603
28866
303
1150
9811
37247
344
1307
14237
54052
415
1574
18714
71047
487
1847
23267
88334
605
2297
–Line Sizes 6- through 8-in.
8-in./DN 200
Rosemount 8600D
kg/hr
lb/hr
5903
22534
528
2015
7747
29575
605
2308
12255
46787
760
2903
21049
80356
996
3804
29761
113440
1184
4520
38342
146375
1345
5134
55640
212411
1620
6185
73135
279200
1901
7259
90931
347134
2364
9025
Reference Manual
00809-0100-4860, Rev BF
kg/hr
2601
233
3414
267
5400
335
9275
439
13093
522
16895
593
24517
714
32226
838
40068
1042
kg/hr
10221
914
13415
1047
21222
1317
36449
1725
51455
2050
66395
2329
96348
2805
126643
3293
157457