Quick Start Guide
2

Wire

2.1
General wiring information
The transmitter is easy to wire.
The front panel is hinged at the bottom. The panel swings down for easy
access to the wiring locations.
2.1.1
Removable connectors and signal input boards
The transmitter uses removable signal input boards and communication
boards for ease of wiring and installation.
You can remove each of the signal boards either partially or completely from
the enclosure for wiring. The transmitter has three slots for placement of up
to two signal input boards and one communication board.
Slot 1 - left
®
Profibus
2.1.2
Signal input boards
Slots 2 and 3 are for signal input measurement boards.
Procedure
1. Wire the sensor leads to the measurement board following the lead
locations marked on the board.
2. Carefully slide the wired board fully into the enclosure slot and take
up the excess sensor cable through the cable gland.
3. Tighten the cable gland nut to secure the cable and ensure a sealed
enclosure.
Note
For the purpose of replacing factory-installed signal input boards,
Rosemount is the sole supplier.
2.1.3
Digital communication boards
®
HART
digital communications is standard on this system. A Profibus
communication board is available as an option for the system's
communication with a host. HART communication supports Bell 202 digital
communications over an analog 4-20 mA current output. Profibus DP is an
open communications protocol which operates over a dedicated digital line
to the host.
22
Slot 2 - center
board
Signal board 1
April 2020
Slot 3 - right
Signal board 2
®
Emerson.com/Rosemount
DP