Quick Start Guide
Tables 1A to 1C, and the Ci and Li of the sensor and interconnected
wiring must be ≤ Ca and La listed in tables 1A to 1C or be classified as
"simple apparatus." Simple apparatus are devices which are incapable of
generating or storing more than 1.2 V, 0.1 A, 25 mW, or 20 µJ
(pH/ORP/ISE and amperometric sensors without preamps and contacing
conductivity sensors qualify as simple apparatus.
J. Rosemount 56 hazardous area
Class I, Division 2, Groups A-D, 0 ° to 50 °C
Class II, III, Division 2, Groups E-G
K. Alarm wiring (Vac) (optional)
L. Analog output (optional)
M. Power supply
N. You may make a ground connection in a hazardous area.
Note
A. Installation must conform to the CEC.
B. Seal required at each conduit entrance.
C. Unless otherwise specified.
20
April 2020
Emerson.com/Rosemount