Install - Emerson Rosemount Clarity II T56 Quick Start Manual

Turbidmeter
Hide thumbs Also See for Rosemount Clarity II T56:

Advertisement

April 2020
1

Install

1.1
Unpack and inspect
Consult the table to verify that you have received the parts for the option
you ordered.
Table 1-1: Rosemount Clarity II Turbidmeter Parts
Item
Single input turbidity transmitter with
®
HART
Dual input turbidity transmitter with
HART
Sensor - EPA standards
Sensor - ISO standard
Cable - 3 ft. (0.9 m)
Cable - 20 ft. (6.1 m)
Cable - 50 ft. (15.2 m)
Calibration cup
Molded chamber/debubbler
1.2
General installation information
1. Although the transmitter is suitable for outdoor use, do not install it
in direct sunlight or in areas of extreme temperatures.
2. Install the transmitter in an area where vibration and
electromagnetic and radio frequency interference are minimized or
absent.
3. Keep the transmitter and sensor wiring at least one foot (0.3 m) from
high voltage conductors. Be sure there is easy access to the
transmitter.
4. The transmitter is suitable for panel, pipe, or surface mounting. Refer
to the figures below.
Quick Start Guide
Model/part number
56-03-27-38-HT
56-03-27-37-HT
8-0108-0002-EPA
8-0108-0003-ISO
2413800
2409700
2409800
2410100
24170-00
Quick Start Guide
5