Wire - Emerson Rosemount Xi Quick Start Manual

Advanced electronics for zirconium oxide flue gas o2 probes
Hide thumbs Also See for Rosemount Xi:

Advertisement

October 2022
2

Wire

2.1
Electrical installation
All wiring must conform to local and national codes. Multiple wiring
diagrams are shown in this section. Always refer to the diagrams that apply
to your transmitter configuration and disregard all other wiring diagrams.
WARNING
ELECTRIC SHOCK
Disconnect and lock out power before connecting the power supply.
Failure to lock out power could result in serious injury or death.
WARNING
ELECTRIC SHOCK
Install all protective covers and safety ground leads after installation.
Failure to install covers and ground leads could result in serious injury or
death.
WARNING
ELECTRIC SHOCK
To meet the safety requirements of EN 61010 (EU requirement) and
ensure safe operation of this equipment, connection to the main electrical
power supply must be made through a circuit breaker (min 10 A) which
will disconnect all current-carrying conductors during a fault situation. This
circuit breaker should also include a mechanically operated isolating
switch. If it does not, locate another external means of disconnecting the
power supply close by. Circuit breakers or switches must comply with a
recognized standard such as IEC 947.
Note
Line, voltage, signal, and relay wiring must be rated for at least 221 °F (105
°C).
NOTICE
If a metal conduit is used with the Rosemount Xi, the conduit should be
reliably bonded to protective earth. The grounding plate inside the
Rosemount Xi is not bounded to PE and does not provide adequate
grounding.
1. Remove cover screws from the front cover of the Rosemount Xi.
Swing down the front cover of the interface box.
Quick Start Guide
Quick Start Guide
17