Install Rosemount 56 transmitter
Figure 1-11: Non-Incendive Field Wiring Installation for the Rosemount
56-27-37 Transmitter
A. Metal conduit
B. If sensor is equipped with a VP connector, the VP connector must be in
the unclassified area.
C. Sensor cable is shielded.
During installation, leave maximum amount of jacket insulation possible
on N.I. field wiring within instrument enclosure. After termination, wrap
N.I. field wiring within enclosure with mylar tape to ensure adequate
double insulation remains.
Max cable length is 50 ft. (15.2 m).
D. Recommended sensor attached. See drawing 1700703 for sensors
recommended for use with options: -30, -31, -32, -33, -34, -35, and -36.
(No sensor if blank or -38).
E.
Use with non-flammable process media only.
F. Rosemount Clarity II turbidity sensor #2 (optional): Option-37
G. Rosemount Clarity II turbidity sensor #1: Option-27
H. Sensor 1. See Drawing 1700703 for sensors recommended for use with
options: -20, -21, -22, -24, -25, and -26.
I. If sensor is equipped with a VP connector, the VP connector must be in
the unclassified area.
J. Rosemount 56 hazardous area
Quick Start Guide