Quick Start Guide
the cable receptacle and the connector on the back of the sensor are
dry when connecting or disconnecting the cable.
8. Place the sensor in either the measuring chamber or the calibration
cup.
Important
The sensor must be in a dark place when power is first applied to
transmitter.
Note
If
confirm sample water flow at debubbler drain outlet.
Important
When using EPA/incandescent sensors (P 8-0108-0002-EPA):
Do not power up the instrument without the sensor connected.
Do not disconnect and reconnect a sensor while a transmitter is
powered.
If this is inconvenient or cannot be avoided:
Cycle power to the instrument after connecting to the sensor or
Perform a slope calibration or standard calibration routine after
connecting the sensor.
Following these guidelines will extend the life of the incandescent lamp and
avoid premature warnings and faults due to reduced lamp life.
26
appears, check sensor cable connection and
S1 Warning
April 2020