Figure 2-27. Event Configuration - Teledyne N500 User Manual

Caps nox analyzer
Table of Contents

Advertisement

this box to "turn off" or deactivate the Event without deleting it). An Event must be
enabled in order to use the Visible and the Latching options.
page when it is triggered (it will still be recorded and can be viewed in the
Utilities>Alerts Log).To use this option, the Event must be enabled.
conditions that triggered it were to correct themselves. (Latching requires that the
user interact with the Active Alerts screen to manually clear the Alert and internal
Event state. Non-latching allows the entry in the Active Alerts screen and the
internal Event state to continuously update based on the Event criteria, requiring no
user interaction to clear the Alert or Event state).
089770000 Rev 05

Figure 2-27. Event Configuration

allows the choice of whether to track and record the Event (uncheck
allows the choice of whether or not to display the Event in the Alerts
allows the choice of whether or not to keep an Event visible even if the
Teledyne API N500 CAPS NO
Analyzer with NumaView™ Software
X
43