Connecting A Vacuum Degasser - Agilent Technologies 1260 Infinity User Manual

Isocratic pump and quaternary pump
Hide thumbs Also See for 1260 Infinity:
Table of Contents

Advertisement

3
Installing the Pump
Connecting Modules and Control Software

Connecting a Vacuum Degasser

The quaternary pump has a built-in degasser. For the isocratic pump, an external degasser
N O T E
can be used and a pump upgrade to the quaterary pump including a built-in degasser is
possible.
1 Place the vacuum degasser in the stack of modules as shown in
2 Connect the bottle head assembly in the solvent reservoir to the degasser
3 Ensure the power switch at the front of the vacuum degasser is OFF (switch
4 Plug an APG cable into the APG remote connector at the rear of the
5 Connect the APG cable to the APG remote connector of the pump, see
6 Press in the power switch to turn on the vacuum degasser.
The AUX output is intended for troubleshooting. It provides a DC voltage in the range of 0 –
N O T E
1 V which is proportional to the vacuum level in the degasser chambers.
54
page 48.
inlet. Connect the degasser outlet to the inlet valve of the pump.
stands out).
degasser.
Figure 11
on page 49.
Agilent 1260 IsoPump/QuatPump User Manual
Figure 10
on