metal housing
polycarbonate
housing
EE22 with fixed probes
position 1
RH probe
∅13
(0.51")
EE22 with remote sensing probe:
∅13
(0.51")
position 2
T probe
∅6
(0.24")
∅12
(0.47")
position 1
RH/T probe
3.

INSTALLATION

3.1

Mounting of metal and polycarbonate housing

1. Drill the mounting holes according to the mounting template.
For the polycarbonte housing a special mounting set for
installation onto mounting rails is available (see chapter 8 "Replacement
parts / Accessories").
2. The bottom part of the housing is mounted with 4 screws (screw
diameter: < 4.2mm
(0.2")
3. Connection of the transmitter (see chapter 4 "Electrical connections").
4. Mounting of the middle part and cover with 4 screws (included in
the scope of supply).
3.2

Mounting of probes

General:
When mounting the probes please note that the RH probe or
RH/T probe has to be mounted always at position 1, the
T probe at position 2. (independent if it is a remote
sensing probe or a fixed probe)
Mounting of fixed sensing probes:
The transmitter must be mounted with the sensing probe
pointed downwards.
Mounting of remote sensing probes:
The sensing probes can be connected to the EE22 housing with
special cables (2, 5 or 10m /
Cable length
2m
(6.6ft)
5m
(16.4ft)
10m
(32.8ft)
The sensor probe must be mounted horizontally or vertically,
pointing downwards. If possible, a dripping water protector
should be fitted.
13
; not included in the scope of supply).
).
6.6ft, 16.4ft or 32.8ft
HA 010801